(Whole Blood/Serum/Plasma) Package Insert REF FIL01 For professional in vitro diagnostic use only. [INTENTED USE] The Filariasis IgG/IgM Rapid Test Cassette (Whole Blood/Serum/Plasma) is a rapid chromatographic immunoassay for the qualitative detection of IgG and IgM antibodies to Filariasis parasites (W. Bancrofti and B. Malayi) in whole blood, serum, or plasma to aid in the diagnosis of Filariasis infection. [SUMMARY] The lymphatic filariasis known as Elephantiasis, mainly caused by W. bancrofti and B. malayi, affects about 120 million people over 80 countries1,2. The disease is transmitted to humans by the bites of infected mosquitoes within which the microflariae sucked from an infected human subject develops into third-stage larvae. Generally, repeated and prolonged exposure to infected larvae is required for establishment of human infection. The definitive parasitologic diagnosis is the demonstration of microflariae in blood samples3. However, this gold standard test is restricted by the requirement for nocturnal blood collection and lack of adequate sensitivity. Detection of circulating antigens is commercially available. Its usefulness is limited for W. bancrofti4. In addition, microfilaremia and antigenemia develop from months to years after exposure. Antibody detection provides an early means to detect filarial parasite infection. Presence of IgM to the parasite antigens suggest current infection, whereas, IgG corresponds to late stage of infection or past infection5. Furthermore, identification of conserved antigens allows 'pan- filaria' test to be applicable. Utilization of recombinant proteins eliminates cross-reaction with individuals having other parasitic diseases6. The Filariasis Rapid Test uses conserved recombinant antigens to simultaneously detect antibody to the W. bancrofti and B. malayi parasites without the restriction on specimen collection. [PRINCIPLE] The Filariasis IgG/IgM Rapid Test Cassette (Whole Blood/Serum/Plasma) is a qualitative membrane based immunoassay for the detection of IgG and IgM antibodies to Filariasis in whole blood, serum, or plasma. In this test procedure, anti-human IgG is immobilized in the IgG line region of the test, anti-human IgM is immobilized in the IgM line region of the test. After specimen is added to the specimen well of the device, it reacts with Filariasis antigen coated particles in the test. This mixture migrates chromatographically along the length of the test and interacts with the immobilized anti-human IgG/ anti-human IgM in respective line. If the specimen contains Filariasis antibodies, a coloured line will appear indicating a positive result. If the specimen does not contain Filariasis antibodies, a coloured line will not appear in IgG/IgM region indicating a negative result. To serve as a procedural control, a coloured line will always appear in the control line region, indicating that proper volume of specimen has been added and membrane wicking has occurred. [REAGENTS] The test contains Filariasis antigen coated particles and anti-human IgG/IgM coated on the IgG/IgM region of membrane. [PRECAUTIONS] |
| • | For professional in vitro diagnostic use only. Do not use after the expiration date. | |
| • | Do not eat, drink or smoke in the area where the specimens or kits are handled. | |
| • | Do not use test if pouch is damaged. | |
| • | Handle all specimens as if they contain infectious agents. Observe established precautions against microbiological hazards throughout testing and follow the standard procedures for proper disposal of specimens. | |
| • | Wear protective clothing such as laboratory coats, disposable gloves and eye protection when specimens are being tested. | |
| • | The used test should be discarded according to local regulations. | |
| • | Humidity and temperature can adversely affect results. |
|
| [STORAGE AND STABILITY] Store as packaged in the sealed pouch at room temperature or refrigerated (2-40°C). The test is stable through the expiration date printed on the sealed pouch. The test must remain in the sealed pouch until use. DO NOT FREEZE. Do not use beyond the expiration date. [SPECIMEN COLLECTION AND PREPARATION] |
||
| • | The Filariasis IgG/IgM Rapid Test Cassette (Whole Blood/Serum/Plasma) can be performed using whole blood (from venipuncture or fingerstick), serum or plasma. | ||
| • | To collect Fingerstick Whole Blood specimens: | ||
| ⚬ | Wash the patient's hand with soap and warm water or clean with an alcohol swab. Allow to dry. | ||
| ⚬ | Massage the hand without touching the puncture site by rubbing down the hand towards the fingertip of the middle or ring finger. | ||
| ⚬ | Puncture the skin with a sterile lancet. Wipe away the first sign of blood. | ||
| ⚬ | Gently rub the hand from wrist to palm to finger to form a rounded drop of blood over the puncture site. | ||
| ⚬ | Add the Fingerstick Whole Blood specimen to the test by using the dropper: | ||
| * | Touch the end of the dropper to the blood and draw up approximately 40 µL. Avoid air bubbles. | ||
| * | Then squeeze the bulb to dispense the whole blood to the specimen area of the test cassette. | ||
| • | Separate the serum or plasma from blood as soon as possible to avoid hemolysis. Only clear, non-hemolyzed specimens can be used. | ||
| • | Testing should be performed immediately after the specimens have been collected. Do not leave the specimens at room temperature for prolonged periods. Serum and plasma specimens may be stored at 2-8°C for up to 3 days. For long term storage, specimens should be kept below 20°C. Whole blood collected by venipuncture should be stored at 2-8°C if the test is to be run within 2 days of collection. Do not freeze whole blood specimens. Whole blood collected by fingerstick should be tested immediately. | ||
| • | Bring specimens to room temperature prior to testing. Frozen specimens must be completely thawed and mixed well prior to testing. Specimens should not be frozen and thawed repeatedly. | ||
| • | If specimens are to be shipped, they should be packed in compliance
with regulations for transportation of etiologic agents. |
||
| [MATERIALS] |
|||
| • Test cassettes • Lancets |
• Droppers • Alcohol Swabs |
• Buffer | • Package insert |
| • Specimen collection containers | • Centrifuge | ||
| • Time | |||
[DIRECTIONS FOR USE] Allow the test, specimen, buffer and/or controls to reach room temperature (15 -30°C) prior to testing. |
||
| 1. | Bring the pouch to room temperature before opening it. Remove the test cassette from the sealed pouch and use it as soon as possible. Best results will be obtained if the assay is performed within one hour. | |
| 2. | Place the cassette on a clean and level surface. For Serum or Plasma specimen: Hold the dropper vertically and transfer 1 drop of serum or plasma (approximately 40µL) to the specimen area,then add 2 drops of buffer (approximately 80 µL),and start the timer, see illustration below. For Venipuncture or Whole Blood specimen: Hold the dropper vertically and transfer 1 drop of whole blood (approximately 40µL) to the specimen area, then add 2 drops of buffer (approximately 80µL), and start the timer. See illustration below. |
|
| 3. | Wait for the coloured line(s) to appear. Read the result at 15 minutes, do not interpret the result after 20 minutes. | |
![]() [INTERPRETATION OF RESULTS] IgM Positive: A coloured line appears in control region (C), another coloured line appears in IgM region. It indicates a IgM positive test result for antibodies to Filariasis. IgG Positive: A coloured line appears in control region (C), another coloured line appears in IgG region. It indicates a IgG positive test result for antibodies to Filariasis. *NOTE: The intensity of the colour in the test line regions (IgM and IgG) may vary depending on the concentration of Filariasis antibodies present in the specimen. Therefore, any shade of colour in the test line region (IgM and/or IgG) should be considered positive. NEGATIVE: One coloured line appears in the control region (C). No apparent coloured line appears in the test region (T). INVALID: Control line fails to appear. Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for control line failure. Review the procedure and repeat the test with a new test cassette. If the problem persists, discontinue using the test kit immediately and contact your local distributor. [QUALITY CONTROL] A procedural control is included in the test. A coloured line appearing in the control region (C) is the internal procedural control. It confirms sufficient specimen volume and correct procedural technique. Control standards are not supplied with this kit; however, it is recommended that positive and negative controls be tested as a good laboratory practice to confirm the test procedure and to verify proper test performance. [LIMITATIONS] |
| 1. | The Assay Procedure and the Test Result Interpretation must be followed closely when testing the presence of antibodies to filarial parasites in serum, plasma or whole blood from individual subjects. Failure to follow the procedure may give inaccurate results. |
| 2. | The Filariasis Rapid Test is limited to the qualitative detection of antibodies to W. bancrofti and B. malayi in human serum, plasma or whole blood. The intensity of the test band does not have linear correlation with the antibody titer in the specimen. |
| 3. | A negative result for an individual subject indicates absence of detectable W. bancrofti and B. malayi antibodies. However, a negative test result does not preclude the possibility of exposure to W. bancrofti and B. malayi. |
| 4. | A negative result can occur if the quantity of W. bancrofti and B. malayi antibodies present in the specimen is below the detection limits of the assay, or the antibodies that are detected are not present during the stage of disease in which a sample is collected. |
| 5. | Some specimens containing unusually high titer of heterophile antibodies or rheumatoid factor may affect expected results. |
| 6. | The results obtained with this test should only be interpreted in conjunction with other diagnostic procedures and clinical findings. |
[EXPECTED VALUES] The Filariasis IgG/IgM Rapid Test Cassette (Whole Blood/Serum/Plasma) has been compared with Culture/Histology, demonstrating an overall accuracy of 97.0%. [PERFORMANCE CHARACTERISTICS] 40 samples from patients with chronic lymphatic filariasis and 251 samples collected from a non-filariasis region were tested by the Filariasis IgG/IgM Rapid Test. Comparison for all subjects is showed in the following table: ![]() Relative Sensitivity: 92.5% (95%CI*: 79.6%-98.4%) *Confidence Interval Relatively Specificity: 98.4% (95%CI*: 96.0%-99.6%) Accuracy: 97.6% (95%CI*: 95.1%-99.0%) IgM Results 32 samples from patients with chronic lymphatic filariasis and 251 samples collected from a non-filariasis region were tested by the Filariasis IgG/IgM Rapid Test. Comparison for all subjects is showed in the following table: ![]() Relative Sensitivity: 96.9% (95%CI*: 83.8%-99.9%) *Confidence Interval Relatively Specificity: 98.8% (95%CI*: 96.5%-99.8%) Accuracy: 98.6% (95%CI*: 96.4%-99.6%) Intra-Assay [BIBLIOGRAPHY] |
| 1. | Lymphatic filariasis: the disease and its control. Fifth report of the WHO Expert Committee on Filariasis. WHO Tech Rep Ser 1992; 281-871. |
| 2. | Michael E, Bundy DAP, Grentell BT. Re-assessing the globalprevalence and distribution of lymphatic filariasis. Parasitology 1996: 112:405-428. |
| 3. | Eberhard ML, Lammie PJ. Laboratory diagnosis of filariasis. Clin. Lab Med 1991; 11:977-1010. |
| 4. | More SJ, Copeman DB. A highly specific and sensitive monoclonal antibody-based ELISA for the detection of circulating antigen in bancroftian filariasis. Trop Med Parasitol 1990; 41:403-406 |
| 5. | Lammie PJ, Weil G, et al: Recombinant antigen-based antibody assays for the diagnosis surveillance of lymphatic filariasis-a multiplecenter trial. Flaria Jornal 2004: 3: 9-18. |
| 6. | Baskar LK, Srikanth TR, et al: Development and evaluation of a rapid flow-through immunofiltration test using recombinant filarial antigen for diagnosis of brugian and bancroftian filariasis. Microbiol Immunol. 2004: 48: 519-25. |
![]() |
|